The MACView® Ethylene Postharvest Portable Analyser is an extreme sensitive ethylene measurement device suitable for measurement of ethylene in postharvest storage rooms, and scrubbers. Levels around 1 ppb resolution can be measured.
Postharvest monitoring
Ethylene (ISO-name ethene), is a plant hormone that is produced by trees, plants, vegetables, fruit, flowers and tulip bulbs. Several cultivars of these vegetations are extremely sensitive for the presence of ethylene in air. Ethylene cause a trigger of ripening and is thus a stress-hormone that can cause irreversible damage to the stored product. Monitoring of ethylene gives information about ethylene production. This information can prevent unexpected rising of ethylene levels that are higher then the harmful level that a product can handle. The MACView® Ethylene Postharvest Portable Analyser is a helpful instrument that helps interpreting what has happened during the storage of your valuable product. Applications The markets that have direct advantage of the MACView® Ethylene Postharvest Analyser are fruit producers (growers), storage facilities, distributors, flower growers, greenhouse owners and research laboratories.
How to use
Most customers that use the MACView® Ethylene Postharvest Portable Analyser says that the value of the instrument is that you can see what happens with the product and that you got the feeling with the process. You will see things that you have never seen before and can relate the change in quality with situations that occur . The portable analyser can be used in any place or circumstances you like. It has a probe that can be inserted in a CA or ULO cell. After the measurement is started up, the measurement will be done in 5 minutes. With a simple press on a button a measurement is started. If necessary the analyser can run in continuous mode. So every 10 minutes, samples are taken and stored in the internal memory. This portable ethylene measurement let you know where ethylene is possible threaten your quality process. For example in storage cells for fruit, the ethylene can be reduced by scrubbing out the ethylene by filters. It gives on long term a better quality of fruit. In several flower types very low levels of around 50 ppb will damage the flowers. For example, roses or phalaenopsis are very sensitive, and any increase of ethylene must be avoided.
Control panel
The MACView® Ethylene Postharvest Portable Analyser is available in 2 measurement ranges: 0-5000 ppb and 0-500 ppm. In the 0-2000 ppb version it gives the measured values in number of ppb’s (parts per billion) on the display. In the 0-500 ppm version it gives a minimum value of 0.1 ppm resolution. Data is logged in the analyser , where also date and time is stored. The complete menu has a intuitive control panel. Connection to external control systems (CA / ULO) This portabLe system is not suitable for a connection with external control systems. For this purpose we have a separate analyser which could be integrated for 24 hours continuous work. In most storage facilities there are control systems for Ultra Low Oxygen (ULO), Controlled Atmosphere (CA), ventilation or circulation systems. These control systems already measure oxygen (O2), and carbon dioxide (CO2). The MACView® Ethylene Postharvest Portable Analyser has an extended range of possibilities for interconnection with these systems. In short can the control system send some pulses for: Start ethylene measurement, start zero measurement and start calibration. The analyser is then functioning as a slave machine and send back the measured values by the analog outputs.
Data / software
With the MACView® Ethylene Postharvest Portable Analyser data ca be logged with date and time stamp. Approximately 6700 record can be logged and the internal rechargeable Lithium-Ion batter pack will work for 16 hours continuously. PC-software suitable fo Windows 95, 98, Windows XP , Windows 2000 and Vista is enclose with the analyser . The analyser and software are available in th languages: English, Dutch, German, Italian, Spanish and French This gives a state of the art control to the system ready for immediate start!
Calibration
The MACView® Ethylene Postharvest Portable Analyser has an extern option for an external low pressure gas bottle that contains calibration gas. The calibration interval can be determined by the user. With a calibration interval of 2 weeks, the calibration bottle can be used approximately 1 year.
Tests carried out by the WUR
The MACView® Ethylene Postharvest Portable Analyser is tested at important institutes in the Netherlands: The device is tested at the WUR in the Netherlands. (WUR = Wageningen University and Research). The WUR department PPO flower bulbs and WUR department Agrotechnology Food Sciences Group (Postharvest Quality) tested the device. They tested our device against their GC and they concluded that the accuracy was better then the lowest accuracy they achieve with their existing GC. (Better then 10 ppb) The WUR also was looking for a good device to do research in practice at farmers locations for the measurement of very low ppb levels. It shows excellent results. The cross sensitivity is none to all common gases in normal air . For example apples emit very much aromates. None of these these aromates does have any effect on the measurement. Also variations like temperature, humidity, CO2 and O2 variations are tested and show all very good results. With many thanks for all the tests and co-operation with the Wageningen University and Research (WUR) Netherlands.
Technical Specifications
Type of instrument | MACView®-Ethylene Postharvest Analyser, based on the solid state electrochemical sensor technology with very high sensitivity and stability |
Versions available | 0-5000 ppb ethylene resolution 1 ppb max. inaccuracy ±0.3‰ 0-500 ppm resolution 0.05 ppm max. inaccuracy ±0.3‰ |
Versions available | Portable version for handheld monitoring Fixed 19” version for fixed mounting to measure rooms of storages |
Option for add a 2nd sensor | Optical O2 sensor measurement range 0 - 25% , 0.001% resolution < 1% O2 |
Option to add a 3th sensor | Optical CO2 sensor measurement range 0 - 10% / 0 - 25% CO2 inaccuracy ±0.1% |
Syringe injection port for MAP | Volume injection for MAP applications, analysis possible by syringe injection |
Software | Standard Web Interface internet connection, either GPRS and UMTS or either Ethernet (Choose which version at purchase) |
Sample speed | 2 modes Real time: 1 sample/sec. / Automatic: 180-7200 seconds per measurement |
Inputs / outputs flow | Combined Input1 / zero input / calibration input (19” also Input 2 available) |
Material of housing | Stainless steel IP62 housing, fixed analyser: is 19” rack mount housing |
Standards | NEN-EN-IEC 61000-6-1 up to 4, CE |
Operation modes | Measurements, dosing, standby, flushing or real-time measurement mode |
Signals / alarms | ppb / ppm hysteresis adjustable on relays (programmable per function), status messages, power, failure |
Data collection | Web enabled / SD-Card internal database, date time for 100.000+ records |
Service connection | USB, SD Card, Serial RS 232 interface and RS485 interface |
Battery Pack | Lithium-Ion 11.500 mAh,charged while system is running, approx. 16 hrs runtime |
Charger | External charger |
Supply | 110 - 230 VAC 75W |
Working temperature | -10 + 50 degrees Celsius, relative humidity 5 - 99%, not condensed |
Languages | Give up 2 languages at order: English, Dutch, German, Italian, Spanish, French |
Control panel | Intuitive menu with graphical display and backlight |
Dimensions | Portable: Width 230 x Depth 405 x Height 180 mm 19” fixed version: Width 480 x Depth 350 x Height 301 mm |
Weight | Portable version: 11.3 Kg Fixed 19” version: 12.8 kg |
Options | Portable version: External calibration package for manual periodic calibration Fixed version: Internal calibration bottle included in the device |